Your Article Summary
Comparison of Three Different Optical Coherence Tomography Models for Total Macular Thickness Measurements in Healthy Controls
Ophthalmologica, 07/02/09
Menke MN et al. - Differences in mean retinal thickness measurements between OCT models were statistically significant. Mean retinal thickness measurement with Cirrus OCT, RTVue OCT and with Stratus OCT was 300, 265 and 257 µm, respectively. Measurements with different OCT models lead to significantly different retinal thickness values.
Marcel N. Menke, 07/02/09
| Recently an increasing number of various OCT models have been introduced. In the future, clinicians as well as researcher will be confronted with OCT data of different OCT models. Therefore it is important to know the differences between these machines. Mainly differences between the automated retinal thickness measurement algorithms account for the variations in thickness measurements. Correction factors may have to be applied to make retinal thickness measurements of different OCT models more comparable. |
Related Articles
Sectoral thinning of the retina after branch retinal artery occlusion
Japanese Journal of Ophthalmology, 11/10/09
Relevance Score: 91%
Oxygen supply and consumption in the retina: implications for studies of retinopathy of prematurity
Documenta Ophthalmologica, 10/20/09
Relevance Score: 90%
The effect of contour line position on optic nerve head analysis by Heidelberg Retina Tomograph
European Journal of Ophthalmology, 11/09/09
Relevance Score: 87%
Combined Three-Dimensional Spectral OCT/SLO Topography and Microperimetry: Steps toward Achieving Functional Spectral OCT/SLO
Ophthalmic Research, 10/16/09
Relevance Score: 87%
Unintentional Displacement of the Retina after Standard Vitrectomy for Rhegmatogenous Retinal Detachment
Ophthalmology, 11/09/09
Relevance Score: 85%
Today in Ophthalmology...keeping you current
Receive free subspecialty "5-minute updates" via email
Comparison Between the plusoptiX and MTI Photoscreeners
Archives of Ophthalmology, 12/18/09
Contact Lens-Related Adverse Events and the Silicone Hydrogel Lenses and Daily Wear Care System Used
Archives of Ophthalmology, 12/18/09
Melanoma-Associated Retinopathy
Archives of Ophthalmology, 12/18/09

See Latest Articles