Lyon ME et al. – All evaluated glucose meter systems demonstrated varying degrees of interference by hematocrit, ascorbate, and maltose mixtures. PQQ–glucose dehydrogenase–based technologies showed greater susceptibility than glucose oxidase–based systems. However, the modified glucose oxidase–based amperometric method was less affected in comparison with the glucose oxidase–based photometric method.