Very low-activity stress/high-activity rest, single-day myocardial perfusion SPECT with a conventional sodium iodide camera and wide beam reconstruction processing
Journal of Nuclear Cardiology, 07/16/2012
DePuey EG et al. – An approximately 185 MBq (5 mCi S) Tc–99m SPECT processed with WBR provides adequate image quality. For larger patients prolonging image acquisition to 16 minutes is beneficial. For patients with normal S SPECT, a S–only protocol is feasible, affording them a very low (approximately 1.4 mSv) radiation dose. If subsequent R SPECT is necessary, it can be performed with approximately 1,332 MBq (36 mCi) with minimal S–R “shine–through.”