Accuracy of a Novel System for Oxygen Delivery to Small Children
Pediatrics, 08/12/2011
Coghill M et al. – Entrainment devices can deliver accurate oxygen concentrations at low flow rates if used with low resistance delivery systems; however, entrainment devices are not suitable for use with high resistance delivery systems such as a standard nasal cannula.
Methods- A continuously adjustable entrainment device was tested to determine the concentrations and flows of oxygen delivered by using low inlet flow rates suitable for therapy for infants and small children and 3 distinct oxygen delivery systems that varied in their resistive load.
- The use of long and large bore, low resistance tubing (similar to a mask) resulted in the delivery of oxygen concentrations that tracked closely (accurate and precise) to values indicated by the entrainment device.
- The directly connected system with lower resistance (similar to a hood) produced a similar profile of concentrations and flow rates to the large bore tubing but with even greater accuracy.
- The use of a long and narrow tubing with higher resistance (similar to a cannula) did not deliver accurate oxygen concentrations.
- In fact, this high-resistance system failed to work as intended, and instead of entraining air, a large proportion (sometimes >50%) of the oxygen delivered to the entrainment device was ejected through its vents.