Directly measured mucosal pressures produced by the i-gel and laryngeal mask airway Supreme in paralysed anaesthetised patients
Anaesthesia,

Eschertzhuber S et al. – The findings suggest that the LMA Supreme mucosal pressures were higher in the hypopharynx than in the distal oropharynx and base of the tongue. There were no pressure differences between the locations for the i–gel.

Methods
  • The i-gel and LMA Supreme are extraglottic airway devices with non-inflatable and inflatable cuffs, respectively.
  • The authors hypothesised that directly measured mucosal pressures would differ between these devices in anesthetised paralysed patients.
  • Thirty patients were randomly allocated to receive one of these two devices.
  • Four pressure sensors were attached to all airway devices used to measure mucosal pressure at the base of the tongue, the distal oropharynx, the hypopharynx and the pyriform fossa.

Results
  • At these four places, median (IQR [range]) i-gel mucosal pressures were 8.0 (2.7-10.7 [0-26.7]), 5.0 (2.7-7 [1.0-37.3]), 9.3 (2.7-13.3 [0-22.7] and 8.0 (2.7-10.7 [0-25.3])cmH2O, respectively, and for the LMA Supreme, these were 5.0 (0.5-8.0 [0-33]), 4.0 (1.3-9.3 [0-24]), 10.7 (4-17.3 [0-26.7]) and 8.0 (0-10.7 [0-36])cmH2O, respectively.
  • Mucosal pressures were low and similar for both devices.

Please login or register to follow this author.
► Click here to access PubMed, Publisher and related articles...
<< Previous Article | Next Article >>

    Currently, there are no available articles.

Your Unread Messages in Anesthesiology

See All >> Messages include industry-sponsored communications and special communications from MDLinx

Most Popular Anesthesiology Articles

Indexed Journals in Anesthesiology: Anaesthesia, Anesthesia & Analgesia, Current Anaesthesia and Critical Caremore